Accuracy of Transcutaneous Bilirubin Measurement
in Term Newborns
Shanika Kosarat MD*,
Varangthip Khuwuthyakorn MD**
Affiliation :
* Department of Pediatrics, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Objective : 1) To evaluate the accuracy of transcutaneous bilirubin (TcB) compared with serum bilirubin (TsB) in full-term
infants, 2) To compare the accuracy of TcB reading from two, three, and four measurements and, 3) To compare the accuracy
of TcB measured at the forehead and sternum.
Material and Method: Full term infants who were diagnosed neonatal jaundice and underwent blood tests for TsB were
eligible for enrollment. Infants who were born at gestational age less than 37 weeks, unstable, previously received
phototherapy, or exchange transfusion were excluded. TcB was measured at forehead and sternum two, three, and four times
at each site. TcB measurements were done within 30 minutes before or after blood sampling for TsB.
Results : The authors obtained 294 paired TcB-TsB from 257 full-term infants. TsB ranged from 1.60 to 21.18 mg/dL (mean
11.03, SD 2.73). The correlation coefficients between TcB at forehead and TsB were significant for two, three, and four
measurements (r = 0.812, 0.800, and 0.800 respectively). TcB measured at sternum also had significant correlation with
TsB (r = 0.829, 0.844, and 0.823 for two, three, and four measurements). TcB tended to underestimate TsB. Measurements
at the sternum seemed to have better correlation with TsB than at the forehead.
Conclusion : TcB has good correlation coefficient to TsB. Two measurements have enough accuracy to estimate TsB level
and measurement at the sternum gives better correlation with TsB than at forehead.
Keywords : Neonatal hyperbilirubinemia, Transcutaneous bilirubin, forehead, sternum