Accuracy of Glucose Meters in Measuring Low Blood Glucose Levels

LULIN CHOUBTUM, B.Sc.*, PAT MAHACHOKLERTWATTANA,M.D.**, UMAPORN UDOMSUBPAYAKUL, M.Sc.*, CHA WALIT PREEYASOMBAT, M.D.**

Affiliation : *Research Center, **Department of Pediatrics, Faculty of Medicine, Ramathibodi Hospital, Mahidol University, Bangkok 10400, Thailand.

Abstract
Background : Hypoglycemia is an emergency condition requiring treatment as soon as possible. Therefore, rapid and reliable blood glucose measurements are necessary. There are 2 sys tems of glucose meters (GMs), the reflectance photometer system (RPS) and the electrochemical biosensor system (BSS). GMs are widely used in monitoring blood glucose (BG) in patients with diabetes. BG values measured by GMs have been confirmed to be accurate especially in measuring normal and high BG levels. However, the data on the accuracy of GMs in measuring low BG levels are limited.
Objective : To compare accuracy and reliability of different systems of GMs in the mea surement of low BG values.
Patients and
Method :
Venous and capillary whole blood specimens were collected from patients who were investigated for pituitary dysfunction. The patients underwent an insulin-induced hypoglycemia test by intravenously administering human regular insulin. The low BG level was defined as having venous plasma glucose (PG) of Jess than 60 mg/dl (mean± SD = 36.59 ± 9.19, n = 54). Capillary blood samples were obtained from fingertips. Venous BG (vBG) and capillary BG (cBG) were measured by GMs. Venous PG which considered a reference value was measured by the glucose dehydrogenase method.
Results : The correlation coefficient (r) between vBG measured by GMs-RPS and PG was 0.86 (p<0.001), whereas, that between vBG by GMs-BSS and PG was 0.75 (p<0.001). Similarly, the r between cBG by GMs-RPS and PG was 0.73 (p<0.001), whereas, that between cBG by GMs BSS and PG was 0.69 (pConclusion : In measuring low blood glucose levels, glucose meters using the reflectance photometer system are more accurate than those using the electrochemical biosensor system.

Keywords : Glucose Meter, Reflectance Photometer, Electrochemical Biosensor, Low Blood Glucose, Hypoglycemia


All Articles Download


INFORMATION

Contact info

JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com

JMed Assoc Thai
MEDICAL ASSOCIATION OF THAILAND
ISSN: 0125-2208 (Print),
ISSN: 2408-1981 (Online)
The content of this site is intended for health professionals.

Submissions

» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement

Other

» Journal Sponsorship » Site Map » About this Publishing System

© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.