Cardiac Magnetic Resonance Imaging of Atrial Septal Defect for Transcatheter Closure

KRITVIKROM DURONGPISITKUL, M.D.*, NAH LEE TANG,DCR***, JARUPIM SOONGSWANG, M.D.*, DUANGMANEE LAOHAPRASITIPORN, M.D.*, APICHART NANA, M.D.*, CHARUWAN KANGKAGATE, B.Sc., M.Sc.**

Affiliation : * Division of Pediatric Cardiology, Department of Pediatrics, ** Her Majesty Cardiac Center, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok 10700, Thailand. *** Philips Electronics Singapore Pte, Ltd, 620A Lorong 1 ToaPayoh, Singapore 319762.

Abstract
Background : The location, size of the defect and age of the patient are the major deter mining factors for transcatheter closure of an atrial septal defect (ASD). The precise shape and anatomy surrounding the defect cannot always be understood by the traditional transesophageal (TEE) echocardiographic technique.
Objectives : The authors compared the measurement of ASD size and atrial septal rim using cardiac Magnetic Resonance Imaging (MRI) and TEE to the balloon sizing technique and device size.
Patients and
Method :
Patients having an ASD which met established criteria were selected for evaluation with cardiac MRI and TEE for a closure procedure. Comparison of the ASD imaging and sizing between the different methods was made.
Results : There were 22 patients who had complete transcatheter closure. The mean age and standard deviation of the patients was 33.2±15.1 (8-67) years old. The mean weight of the patients was 51.6±13.1 (20-99) kg. The average cardiac MRI measurement of the ASD was 24.9± 6.4 mm compared to the TEE measurement of 20.8±5.5 mm. The transcatheter balloon measure ment of the ASD was 25.2±6.9 (11-36) mm and the device closure size was 24.8±6.6 (11-36) mm. The correlation coefficient of cardiac MRI to device closure size was r = 0.784 (p < 0.001) when compared to TEE measurement to device closure size; r = 0.761 (p = 0.001).
Conclusion : The authors demonstrated the capability of the cardiac MRI in assessment of the ASD morphology and anatomy for transcatheter closure of the ASD with an Amplatzer™ Septal Occluder. Cardiac MRI can provide information about the type, location, size of the defect and direct visualization of the atrial septum anatomy. This detailed information enabled us to pro vide a safer, more effective application of the ASD occluder.

Keywords : Cardiac Magnetic Resonance Imaging and Atrial Septal Defect


All Articles Download


INFORMATION

Contact info

JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com

JMed Assoc Thai
MEDICAL ASSOCIATION OF THAILAND
ISSN: 0125-2208 (Print),
ISSN: 2408-1981 (Online)
The content of this site is intended for health professionals.

Submissions

» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement

Other

» Journal Sponsorship » Site Map » About this Publishing System

© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.