Purpose :
To compare the measurement of central corneal thickness between ultrasonic and optical techniques in normal Thai myopic eyes.
Method : In this prospective study, the ultrasonic and optical pachymeter were employed to measure the central corneal thickness in 100 eyes of normal myopic volunteers. Corneal thickness was compared using the paired Student t-test. The correlation of central corneal thickness between the two groups was assessed by linear regression analysis.
Results : The difference of mean central corneal thickness between ultrasound (554.4 ± 27.50 J.!m) and optical (581.1 ± 22.62 J.!m) pachymetry was statistically significant in the normal myopic eyes = = = (p 0.00 I). Both methods possessed a highly significant linear correlation (r 0.90, p 0.001)
Conclusions : The optical measurement of central corneal thickness in normal myopic eyes is, on average, 27 J.!m greater than ultrasonic pachymeter measurement.

Keywords : Central Corneal Thickness, Ultrasonic Pachymetry, Optical Pachymetry


All Articles Download


INFORMATION

Contact info

JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com

JMed Assoc Thai
MEDICAL ASSOCIATION OF THAILAND
ISSN: 0125-2208 (Print),
ISSN: 2408-1981 (Online)
The content of this site is intended for health professionals.

Submissions

» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement

Other

» Journal Sponsorship » Site Map » About this Publishing System

© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.