Submit manuscript

Detection of Orthopedics Implants in Vivo by Hand-Held Metal Detectors

Pokbhatara Kunasuntiwarakul, MD¹, Sompob Poopitaya, MD¹

Affiliation : ¹ Department of Orthopedic, Phramongkutklao Hospital, Bangkok, Thailand


Objective: To assess rates of detection of various orthopedic implants by airport detectors with the new security sensitivities.
Materials and Methods: Two hundred sixty-one volunteers with 386 implants were asked to be scanned by a hand-metal detector model Garrett SUPER SCANNER™. The sensitivity was set equivalent to the Suvarnabhumi Airport Transportation Security Administration setting for regular security.
Results: Of the 386 implants in 261 patients who were screened, 231 (60%) were trauma hardware, including intramedullary nails, plates, screws, and Kirschner wires, and 155 (40%) were arthroplasty implants. Three hundred thirty-five (86.78%) of the 386 implants were detected by the metal detector. The overall rate of detection was 99.35% for prosthetic replacements and 98.05% for plates. All the total knee replacements and 95% of the total hip replacements were detected. The Kirschner wires were not detected. The overall detection rate was 92.54% for implants in the lower extremity, 55.73% for those in the upper extremity, and 93.33% for those in the spine. The detection rate for implants in the lower extremity was two times higher than the implants in the upper extremity and equal for implants in the spine.
Conclusion: Most of the orthopedic implants may be detected by hand-metal detector model metal detector used at commercial airports. Total joint prostheses, nails, and plates will routinely set off the detector, whereas screws and wires are rarely detected. Cobalt-chromium and titanium implants are more likely to be detected than stainless-steel implants.
Received 12 February 2018 | Revised 15 April 2018 | Accepted 16 April 2018

doi.org/10.35755/jmedassocthai.2020.09.11921

Keywords : Orthopedic Implant, Hand-Held Metal Detector, Security


All Articles Download


INFORMATION

Contact info

JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com

JMed Assoc Thai
MEDICAL ASSOCIATION OF THAILAND
ISSN: 0125-2208 (Print),
ISSN: 2408-1981 (Online)
The content of this site is intended for health professionals.

Submissions

» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement

Other

» Journal Sponsorship » Site Map » About this Publishing System

© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.