Supaporn Trakanwitthayarak, MD¹
Affiliation : ¹ Department of Ophthalmology, Ratchaburi Hospital, Ratchaburi, Thailand
Objective: To compare the intraocular pressure (IOP) values acquired from the non-contact tonometers and a Goldmann
applanation tonometer in glaucoma patients.
Materials and Methods: The present study included 300 eyes from 150 participants that attended the glaucoma outpatient clinic. The IOP was measured using both non-contact tonometry (NCT) and Goldmann applanation tonometry (GAT). The differences in IOP readings between the two techniques were evaluated.
Results: The mean IOP as measured by NCT was 16.26±6.95 mmHg, when that of measured by GAT was 16.11±8.43 mmHg. The mean difference between the two techniques of measurement was 0.147±3.01 mmHg. The values acquired from NCT were slightly higher than those acquired by GAT in 49% of patients, and this difference was more distinct when the IOP as measured by GAT more than 21 mmHg.
Conclusion: There was no statistically significant correlation in the measurement of IOP between non-contact and GAT tonometers. NCT is a proper method for mass screenings of IOP even if the IOP measurement by NCT is slightly higher than by GAT.
Received 18 Jun 2019 | Revised 16 Sep 2019 | Accepted 18 Sep 2019
doi.org/10.35755/jmedassocthai.2020.08.10279
Keywords : Tonometry, Comparison, Glaucoma, Non-contact tonometry, Goldmann applanation tonometer
JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com
» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement
» Journal Sponsorship » Site Map » About this Publishing System
© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.